Impedance
Probe tone:
226Hz +/-.01% 85dBSPL +/- 1dB
687Hz +/- .01% 80dB SPL +/- 1dB
1kHz +/- .01% 78dB SPL +/- 1dB
EV Range: 226Hz 0.2ml to 5ml
687Hz 0.1ml to 2ml
1kHz 0.1ml to 2ml
Compliance accuracy: +/- 5%
Pressure range: Normal -400 to +200
Extended -600 to +300
pressure accuracy: +/- 5%
Pump speed: Low 100 daPa/s
Medium 200daPa/s
High 300daPa/s
Reflex
Ipsilateral stimulation Frequencies: 250Hz, 500, 1k, 2k, 3k, 4k
Noise: White noise, High pass, Low pass Intensities: 0 - 110dBHL
ContraIateral stimulation Frequencies: 250, 500, 1k, 2k, 3k, 4k,6k, 8kHz
Noise: White noise, High pass, Low pass
Intensities: 0 - 120dBHL
Frequency Accuracy: +/- 1Hz
Harmonic distortion: less than 1%